Comparison of retention forces with various fabrication methods and materials in double crowns

双冠不同制作方法和材料的固位力比较

阅读:1
作者:Melahat Çelik Güven, Meral Tuna, Ergun Bozdağ, Gizem Nur Öztürk, Gulsen Bayraktar

Conclusion

More predictable, clinically relevant, and less excursive retention forces can be observed in the ZA groups. The retention force values of the LL groups were statically similar to those of the other groups, except the ZA groups.

Methods

Ten groups, each consisting of six samples, were evaluated. Specifically, casting gold alloy primary crown - casting gold alloy secondary crown (AA), laser sintering primary crown - laser sintering secondary crown (LL), casting Cr alloy primary crown - casting Cr alloy secondary crown, (CC) zirconia primary crown - electroformed secondary crown (ZA), and CAD/CAM titanium alloy primary crown - CAD/CAM titanium alloy secondary crown (TT) groups were evaluated at cone angles of 4° and 6°. The samples were subjected to 5,000 insertion-separation cycles in artificial saliva, and the retention forces were measured every 500 cycles. The wear levels were analyzed via SEM at the beginning and end of the 5,000 cycles.

Purpose

The purpose of this study was to analyze the retention force changes and wear behaviours of double-crown systems over long-term use. Materials and

Results

In all samples, the retention forces increased when the conus angle decreased. The highest initial and final retention force values were found in the LL-4° group (32.89 N-32.65 N), and the lowest retention force values were found in the ZA6° group (5.41 N-6.27 N). The ZA groups' samples showed the least change in the retention force, and no wear was observed. In the other groups, wear was observed mostly in the primary crowns.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。