Characterization of atherosclerotic plaque: a contrast-detail study using multidetector and cone-beam computed tomography

动脉粥样硬化斑块的特征:使用多探测器和锥形束计算机断层扫描进行对比细节研究

阅读:4
作者:Nima Kasraie, Peter Mah, Carl R Keener, Geoffrey D Clarke

Abstract

A Hindmarsh-Rose model perceptibility phantom containing inserts with various in vitro atherosclerotic plaque compositions was constructed and imaged on a clinical 64 slice multidetector (MDCT) system using 80 and 120 kVp settings and two other cone-beam (CBCT) systems at 80 kVp. Perceptibility of the simulated lipid-fibrotic plaque solutions in the images was evaluated by six observers. The effective doses of the protocols employed were estimated using phantom CTDI-vol measurements placed at identical settings. The CBCT system allowed reduction in effective dose in comparison with the conventional MDCT system for imaging of the carotid plaque phantoms without degrading image quality. The CBCT dose was less than MDCT, with a mean dose of 1.14 ± 0.01 mSv and 1.11 ± 0.02 mSv for MDCT using two measuring techniques vs. 0.35 ± 0.01 mSv for CBCT. The image quality analysis showed no significant differences in the contrast-detail (C-D) curves of the best performing CBCT vs. clinical MDCT (p > 0.05) using a Mann-Whitney U test. Results indicate that low-tube-potential CBCT may produce comparable C-D resolution for phantom-based representations of soft plaque types with respect to MDCT systems. This study suggests that the utility of low kVp CT techniques for evaluating carotid vulnerable atherosclerotic plaque merits further study.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。