Three-dimensional macroporous titanium-aluminum-vanadium (TiAl6V4) implants produced by additive manufacturing (AM) can be grit blasted (GB) to yield microtextured exterior surfaces, with additional micro/nano-scale surface features provided by subsequent acid etching (AE). However, the line-of-sight nature of GB causes the topography of exterior GBâ+âAE-modified surfaces to differ from internal GB-inaccessible surfaces. Previous in vitro studies using dense TiAl6V4 substrates indicated that a nonline-of-sight, calciothermic-reaction (CaR)-based process provided homogeneous osteogenic nanotextures on GBâ+âAE surfaces, suggesting it could be used to achieve a homogeneous nanotopography on external and internal surfaces of macroporous AM constructs. Macroporous TiAl6V4 (3D) constructs were produced by direct laser melting and modified by GBâ+âAE, with the CaR process then applied to 50% of constructs (3DCaR). The CaR process yielded nanoporous/nanorough internal surfaces throughout the macroporous constructs. Skeletally mature, male Sprague-Dawley rats were implanted with these constructs using a cranial on-lay model. Prior to implantation, a Cu++-free click hydrogel was applied to half of the constructs (3Dâ+âH, 3DCaRâ+âH) to act as a challenge to osseointegration. Osseointegration was compared between the four implant groups (3D, 3DCaR, 3Dâ+âH, 3DCaRâ+âH) at 4w. 3Dâ+âH implants exhibited lower bone volume (BV) and percent bone ingrowth (%BI) than the 3D implants. In contrast, osseointegrated 3DCaRâ+âH implants had similar BV and %BI as the 3DCaR implants. Implant pull-off forces correlated with these results. In vitro analyses indicated that human bone marrow stromal cells (MSCs) exhibited enhanced production of osteoblast differentiation markers and factors associated with osteogenesis when grown on CaR-modified 3D substrates relative to control (TCPS) substrates. This work confirms that the CaR process provides osteogenic nanotextures on internal surfaces of macroporous 3D implants, and suggests that CaR-modified surfaces can promote osseointegration in cases where osteogenesis is impaired.
Internal surface modification of additively manufactured macroporous TiAl6V4 biomimetic implants via a calciothermic reaction-based process and osteogenic in vivo responses.
通过基于钙热反应的工艺对增材制造的大孔 TiAl6V4 仿生植入物进行内表面改性,并研究其在体内的成骨反应
阅读:7
作者:Berger Michael B, Cohen D Joshua, Deng Jingyao, Srivas Pavan, Boyan Barbara D, Sandhage Kenneth H, Schwartz Zvi
| 期刊: | Journal of Biomedical Materials Research Part B-Applied Biomaterials | 影响因子: | 3.400 |
| 时间: | 2024 | 起止号: | 2024 Jan;112(1):e35322 |
| doi: | 10.1002/jbm.b.35322 | 研究方向: | 骨科研究 |
特别声明
1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。
2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。
3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。
4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。