In this study, we conducted a numerical analysis on catheter sizes using computational fluid dynamics to assess urinary flow rates during intermittent catheterization (IC). The results revealed that the fluid (urine) movement within a catheter is driven by intravesical pressure, with friction against the catheter walls being the main hindrance to fluid movement. Higher-viscosity fluids experienced increased friction with increasing intravesical pressure, resulting in reduced fluid velocity, whereas lower-viscosity fluids experienced reduced friction under similar pressure, leading to increased fluid velocity. Regarding urine characteristics, the results indicated that bacteriuria, with lower viscosity, exhibited higher flow rates, whereas glucosuria exhibited the lowest flow rates. Additionally, velocity gradients decreased with increasing catheter diameters, reducing friction and enhancing fluid speed, while the friction increased with decreasing diameters, reducing fluid velocity. These findings confirm that flow rates increased with larger catheter sizes. Furthermore, in terms of specific gravity, the results showed that a 12Fr catheter did not meet the ISO-suggested average flow rate (50Â cc/min). The significance of this study lies in its application of fluid dynamics to nursing, examining urinary flow characteristics in catheterization. It is expected to aid nurses in selecting appropriate catheters for intermittent catheterization based on urinary test results.
Analysis of the urine flow characteristics inside catheters for intermittent catheter selection.
分析导尿管内尿流特性,以选择间歇性导尿管
阅读:5
作者:Lee Kyeongeun, Han Jeongwon
| 期刊: | Scientific Reports | 影响因子: | 3.900 |
| 时间: | 2024 | 起止号: | 2024 Jun 10; 14(1):13273 |
| doi: | 10.1038/s41598-024-64395-9 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。