Magnetic Resonance Imaging-Compatible Optically Powered Miniature Wireless Modular Lorentz Force Actuators.

磁共振成像兼容的光供电微型无线模块化洛伦兹力致动器

阅读:4
作者:Mutlu Senol, Yasa Oncay, Erin Onder, Sitti Metin
Minimally invasive medical procedures under magnetic resonance imaging (MRI) guidance have significant clinical promise. However, this potential has not been fully realized yet due to challenges regarding MRI compatibility and miniaturization of active and precise positioning systems inside MRI scanners, i.e., restrictions on ferromagnetic materials and long conductive cables and limited space around the patient for additional instrumentation. Lorentz force-based electromagnetic actuators can overcome these challenges with the help of very high, axial, and uniform magnetic fields (3-7 Tesla) of the scanners. Here, a miniature, MRI-compatible, and optically powered wireless Lorentz force actuator module consisting of a solar cell and a coil with a small volume of 2.5 × 2.5 × 3.0 mm(3) is proposed. Many of such actuator modules can be used to create various wireless active structures for future interventional MRI applications, such as positioning needles, markers, or other medical tools on the skin of a patient. As proof-of-concept prototypes toward such applications, a single actuator module that bends a flexible beam, four modules that rotate around an axis, and six modules that roll as a sphere are demonstrated inside a 7 Tesla preclinical MRI scanner.