Data-independent acquisition (DIA) mass spectrometry facilitates high-throughput, reproducible bottom-up proteomic analyses. Typically, DIA methods coselect multiple precursor ions within a wide selection window. These precursors are simultaneously fragmented, superimposing the product ion signals into a complex chimeric spectrum. A method for varying the quadrupole selection width over the ion accumulation period is described. This method couples the intensity of a product ion to the mass of its precursor ion. By overlapping consecutive selection windows, scan-to-scan product ion intensity profiles can be used to infer precursor mass. We assess the method's sensitivity to quadrupole width, accumulation time, and mass-to-charge range using internal fluoranthene calibrant and FlexMix calibration solution with Q-Orbitrap configured mass analyzers. Additionally, we explore usability of the described technique on a tryptic-digest monoclonal antibody sample, including both direct infusion and liquid chromatography of the sample. With direct infusion, product ions from two precursors separated by 1 thomson (Th) are resolved with this method using 10 Th windows with 5 Th overlap. The product ions are associated within 0.3 Th of their respective precursor ion's m/z. Therefore, product ion spectra have a precursor ion m/z resolving power of â¼33.
Dynamic Quadrupole Selection to Associate Precursor Masses with MS/MS Products in Data-Independent Acquisition.
在数据非依赖采集中,动态四极杆选择将前体质量与 MS/MS 产物关联起来
阅读:9
作者:Mertz Keaton L, Serrano Lia R, Sinitcyn Pavel, Coon Joshua J
| 期刊: | Journal of the American Society for Mass Spectrometry | 影响因子: | 2.700 |
| 时间: | 2025 | 起止号: | 2025 Sep 3; 36(9):1869-1876 |
| doi: | 10.1021/jasms.5c00110 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。