Bioinertization of NanoLC/MS/MS Systems by Depleting Metal Ions From the Mobile Phases for Phosphoproteomics

通过消耗流动相中的金属离子对 NanoLC/MS/MS 系统进行生物惰性化,用于磷酸化蛋白质组学研究

阅读:9
作者:Yumi Komori, Tomoya Niinae, Koshi Imami, Jun Yanagibayashi, Kenichi Yasunaga, Shinya Imamura, Masami Tomita, Yasushi Ishihama

Abstract

We have successfully developed a bioinertized nanoflow LC/MS/MS (nanoLC/MS/MS) system for the highly sensitive analysis of phosphopeptides by depleting metal ions from the mobile phase. We found that not only direct contact of phosphopeptides with metal components, but also indirect contact with nanoLC pumps through the mobile phase causes significant losses during the recovery of phosphopeptides. Moreover, electrospray ionization was adversely affected by the mobile phase containing multiple metal ions as well as by the sample solvents contaminated with metal ions used in immobilized metal ion affinity chromatography for phosphopeptide enrichment. To solve these problems, metal ions were depleted by inserting an online metal ion removal device containing metal-chelating membranes between the gradient mixer and the autosampler. As a result, the peak areas of the identified phosphopeptides increased an average of 9.9-fold overall and 77-fold for multiply phosphorylated peptides with the insertion of the online metal ion removal system. This strategy would be applicable to the highly sensitive analysis of other phosphorylated biomolecules by microscale-LC/MS/MS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。