Traditional Western blots are commonly used to separate and assay proteins; however, they have limitations including a long, cumbersome process and large sample requirements. Here, we describe a system for Western blotting where capillary gel electrophoresis is used to separate sodium dodecyl sulfate-protein complexes. The capillary outlet is threaded into a piezoelectric inkjetting head that deposits the separated proteins in a quasi-continuous stream of <100 pL droplets onto a moving membrane. Through separations at 400 V/cm and protein capture on a membrane moving at 2 mm/min, we are able to detect actin with a limit of detection at 8 pM, or an estimated 5 fg injected. Separation and membrane capture of sample containing 10 proteins ranging in molecular weights from 11 - 250 kDa was achieved in 15 min. The system was demonstrated with Western blots for actin, β-tubulin, ERK1/2, and STAT3 in human A431 epidermoid carcinoma cell lysate.
Capillary electrophoresis Western blot using inkjet transfer to membrane.
采用喷墨转印法将毛细管电泳蛋白印迹法转移到膜上
阅读:13
作者:Booth Peter-Philip M, Lamb Don T, Anderson Jon P, Furtaw Michael D, Kennedy Robert T
| 期刊: | Journal of Chromatography A | 影响因子: | 4.000 |
| 时间: | 2022 | 起止号: | 2022 Aug 30; 1679:463389 |
| doi: | 10.1016/j.chroma.2022.463389 | 方法学: | WB |
| 研究方向: | 其它 | ||