Multicenter evaluation of the QIAstat-Dx Gastrointestinal Panel 2, a multiplex PCR platform for the diagnosis of acute gastroenteritis.

阅读:3
作者:Szymczak Wendy A, Engsbro Anne Line, Lisby Jan Gorm, González-López Juan José, Granato Paul, Ledeboer Nathan, Wolk Donna M, Young Stephen, Rhoads Daniel D, Caspar Yvan, Steed Lisa, Humphries Romney, Bielefeldt Christopher, Hermanowski Markus, de Diego Juana L, Leibhan Hendrik, Boher Pau, Camprubí Carla, Orthodoxou Maria, Sala Ester, Johnson Sarah, Juanola-Falgarona Martí, Manissero Davide, Bialas Johanna
The QIAstat-Dx Gastrointestinal Panel 2 (GI2 Panel) is a sample-to-answer multiplex PCR instrument that can detect 17 targets in a run time of about 80 minutes. The performance of the QIAstat-Dx GI2 Panel was evaluated by testing 1,939 prospective, 119 prospectively collected and then archived positive clinical samples and 750 retrospective clinical specimens across 13 sites in Europe and the United States. Specimens tested included bulk stool samples preserved in modified Cary-Blair transport medium. For most targets, results were compared to those of the FilmArray GI panel (13/17), and discordant results were adjudicated with a third assay. For the remaining targets (4/17), a composite comparator method was used, which included three comparator assays for each target. Before discordant resolution, the QIAstat-Dx GI2 Panel positive percent agreement (PPA) was 95% or greater for 5/17 targets (Campylobacter, E. coli O157, Cryptosporidium, Cyclospora cayetanensis, and Giardia lamblia) and 90% or greater for 11/17 targets: adenovirus F40/F41, astrovirus, norovirus GI/GII, rotavirus A, Plesiomonas shigelloides, enteropathogenic Escherichia coli, enterotoxigenic E. coli, Salmonella, Yersinia enterocolitica, Shiga-like toxin E. coli (STEC) stx1/stx2, and Shigella/enteroinvasive E. coli. No cases of Entamoeba histolytica were encountered during the clinical study. The negative percent agreement (NPA) was >98.9% for all QIAstat-Dx GI2 Panel targets. The three most common pathogens identified in single and co-infections were enteropathogenic E. coli (9.9%), Campylobacter (5.2%), and norovirus GI/GII (3.1%). In summary, this clinical study examined more than 2,800 samples from Europe and the U.S. using the QIAstat-Dx GI2 Panel and identified 90%-100% PPA and 99% NPA for its 17 targets.IMPORTANCEThe manuscript highlights the significance and impact of the QIAstat-Dx GI2 Panel, a sample-to-answer multiplex PCR instrument capable of detecting 17 targets in approximately 80 minutes. This comprehensive clinical study, conducted across 13 sites in Europe and the United States, evaluated the performance of the panel using over 2,800 clinical samples. The results demonstrate a high accuracy of the QIAstat-Dx GI2 panel, with a PPA equal to or higher than 90% for all targets and an NPA greater than 98.9% for all targets. These findings underscore the reliability and effectiveness of the GI2 panel in the rapid and precise detection of gastrointestinal pathogens, which is crucial for timely diagnosis and treatment of infections.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。