Polymeric biomedical implants are an important clinical tool, but degradation remains difficult to determine post-implantation. Computed tomography (CT) could be a powerful tool for device monitoring, but polymers require incorporation of radiopaque contrast agents to be distinguishable from tissue. In addition, immune response to radiopaque devices must be characterized as it modulates device function. Radiopaque devices and films were produced by incorporating 0-20 wt% TaO(x) nanoparticles into polymers: polycaprolactone (PCL) and poly(lactide-co-glycolide) (PLGA). In vitro inflammatory responses of mouse bone marrow-derived macrophages to polymer matrix incorporating TaO(x) nanoparticles was determined by monitoring cytokine secretion. Nanoparticle addition stimulated a slight inflammatory reaction, increasing TNFα secretion, mediated by changes in polymer matrix properties. Subsequently, devices (PLGA 50:50 + 20 wt% TaO(x)) were implanted subcutaneously in a mouse model of chronic inflammation, that featured a sustained increase in inflammatory response local to the implant site over 12 weeks. No changes to device degradation rates or foreign body response were noted between a normal and chronically stimulated inflammatory environment. Serial CT device monitoring post-implantation provided a detailed timeline of device collapse, with no rapid, spontaneous release of nanoparticles that occluded matrix visualization. Importantly, repeat CT sessions did not ablate the immune system or alter degradation kinetics. Thus, polymer devices incorporating radiopaque nanoparticles can be used for in situ monitoring and be readily combined with other medical imaging techniques, for a dynamic view biomaterial and tissue interactions. STATEMENT OF SIGNIFICANCE: A growing number of implantable devices are in use in the clinic, exposing patients to inherent risks of implant movement, collapse, and infection. The ability to monitor implanted devices would enable faster diagnosis of failure and open the door for personalized rehabilitation therapies - both of which could vastly improve patient outcomes. Unfortunately, polymeric materials which make up most biomedical devices are not radiologically distinguishable from tissue post-implantation. The introduction of radiopaque nanoparticles into polymers allows for serial monitoring via computed tomography, without affecting device degradation. Here we demonstrate for the first time that nanoparticles do not undergo burst release from devices post-implantation and that inflammatory responses - a key determinant of device function in vivo - are also unaffected by nanoparticle addition.
In vivo micro-computed tomography evaluation of radiopaque, polymeric device degradation in normal and inflammatory environments.
利用体内微型计算机断层扫描技术评估正常和炎症环境下放射性不透光聚合物装置的降解情况
阅读:12
作者:Pawelec Kendell M, Hix Jeremy M L, Troia Arianna, MacRenaris Keith W, Kiupel Matti, Shapiro Erik M
| 期刊: | Acta Biomaterialia | 影响因子: | 9.600 |
| 时间: | 2024 | 起止号: | 2024 Jun;181:222-234 |
| doi: | 10.1016/j.actbio.2024.04.031 | 研究方向: | 炎症/感染 |
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。