Assessment of hemodynamic parameters by PiCCO and PAC in patients treated with the Impella CP.

通过 PiCCO 和 PAC 对接受 Impella CP 治疗的患者的血流动力学参数进行评估

阅读:4
作者:Schmitt Dominik, Schneider Andreas, Lengenfelder Björn, Wagner Martin, Patschan Daniel, Sasko Benjamin, Ertl Georg, Frantz Stefan, Ritter Oliver, Weismann Dirk
In 28 patients supported by an Impella pump (Impella CP, Abiomed Inc.(®), Danvers, MA), hemodynamic measurements by PiCCO (PULSION Medical Systems SE(®), Feldkirchen, Germany) and pulmonary artery catheter (PAC or Swan-Ganz Catheter; Edwards Lifescience(®), Unterschleissheim, Germany) were compared. There was a significant positive correlation of cardiac output (CO; r (2) = 0.917, p < 0.001), systemic vascular resistance index (SVRI; r (2) = 0.904, p < 0.001), stroke volume index (SI; r (2) = 0.909, p < 0.001) and left ventricular work index (LCWI; r (2) = 0.689, p < 0.001) in PiCCO and PAC measurements under Impella CP support. We conclude, that in patients with left-ventricular Impella support, hemodynamics may be assessentd by PiCCO or PAC.