Highly sensitive and accurate detection of C-reactive protein by CdSe/ZnS quantum dot-based fluorescence-linked immunosorbent assay

基于CdSe/ZnS量子点的荧光免疫吸附分析高灵敏、准确检测C反应蛋白

阅读:9
作者:Yanbing Lv, Ruili Wu, Kunrui Feng, Jinjie Li, Qing Mao, Hang Yuan, Huaibin Shen, Xiangdong Chai, Lin Song Li

Background

The conventional and widely used enzyme-linked immunosorbent assays (ELISA), due to specificity and high-sensitivity, were suitable in vitro diagnosis. But enzymes are vulnerable to the external conditions, and the complex operation steps limit its application. Semiconductor quantum dots have been successfully used in biological and medical research due to the high photoluminescence and high resistance to photobleaching. In this study, we have developed a novel quantum dot-labeled immunosorbent assay for rapid disease detection of C-reactive protein (CRP).

Conclusions

This in vitro quantum dot-based detection method offers a lower LOD and a wide liner detection range than ELISA. The total reaction time is only 50 min, which is much shorter than the commercialization ELISA (about 120 min). All of the results show that a convenient, sensitive, and accurate fluorescence-linked immunosorbent assay method has been well established for the detection of CRP samples. Therefore, this method has immense potential for the development of rapid and cost-effective in vitro diagnostic kits.

Results

The assay for the detection of CRP can provide a wide analytical range of 1.56-400 ng/mL with the limit of detection (LOD) = 0.46 ng/mL and the limit of quantification = 1.53 ng/mL. The precision of the assay has been confirmed for low coefficient of variation, less than 10% (intra-assay) and less than 15% (inter-assay). The accuracy of assay meets the requirements with the recoveries of 95.4-105.7%. Furthermore, clinical samples have been collected and used for correlation analysis between this FLISA and gold standard Roche immunoturbidimetry. It shows excellent accurate concordance and the correlation coefficient value (R) is as high as 0.989 (n = 34). Conclusions: This in vitro quantum dot-based detection method offers a lower LOD and a wide liner detection range than ELISA. The total reaction time is only 50 min, which is much shorter than the commercialization ELISA (about 120 min). All of the results show that a convenient, sensitive, and accurate fluorescence-linked immunosorbent assay method has been well established for the detection of CRP samples. Therefore, this method has immense potential for the development of rapid and cost-effective in vitro diagnostic kits.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。