Electromyostimulation (EMS) is an up-and-coming training method that demands further fundamental research regarding its safety and efficacy. To investigate the influence of different stimulation parameters, electrode positions and electrode sizes on the resulting voltage in the tissue, a tissue mimicking phantom is needed. Therefore, this study describes the fabrication of a hydrogel arm phantom for EMS applications with the tissue layers of skin, fat, blood and muscle. The phantom was dielectrically validated in the frequency range of 20 Hz to 100 Hz. We also conducted electromyography (EMG) recordings during EMS on the phantom and compared them with the same measurements on a human arm. The phantom reproduces the dielectric properties of the tissues with deviations ranging from 0.8% to more than 100%. Although we found it difficult to find a compromise between mimicking the permittivity and electrical conductivity at the same time, the EMS-EMG measurements showed similar waveforms (1.9-9.5% deviation) in the phantom and human. Our research contributes to the field of dielectric tissue phantoms, as it proposes a multilayer arm phantom for EMS applications. Consequently, the phantom can be used for initial EMS investigations, but future research should focus on further improving the dielectric properties.
Fabrication and Dielectric Validation of an Arm Phantom for Electromyostimulation.
阅读:7
作者:Uhrhan Katja, Schwindt Esther, Witte Hartmut
| 期刊: | Bioengineering-Basel | 影响因子: | 3.800 |
| 时间: | 2024 | 起止号: | 2024 Jul 17; 11(7):724 |
| doi: | 10.3390/bioengineering11070724 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。