Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.

阅读:4
作者:Liao Ying-Lan, Lai Nan-Ku, Tyan Yeu-Sheng, Tsai Hui-Yu
BACKGROUND: To quantify image quality and radiation doses in regions adjacent to and distant from bismuth shields in computed tomography (CT). METHODS: An American College of Radiology accreditation phantom with four solid rods embedded in a water-like background was scanned to verify CT number (CTN) accuracy when using bismuth shields. CTNs, image noise, and contrast-to-noise ratios (CNRs) were determined in the phantom at 80-140 kVp. Image quality was investigated on image portions in the zones adjacent (A zone) to and distant (D zone) from a bismuth shield. Surface radiation doses were measured using thermoluminescent dosimeters. Streak artefacts were graded on a 3-point-scale. RESULTS: Changes in CTN caused by a bismuth shield resulted in changes in X-ray spectra. CTN changes were more apparent in the A zone than in the D zone, particularly for a low tube voltage. The degrees of CTN changes and image noise were proportional to the thickness of the bismuth shields. A 1-ply bismuth shield reduced surface radiation doses by 7.2%-15.5%. The overall CNRs were slightly degraded, and streak artefacts were acceptable. CONCLUSIONS: Using a bismuth shield could result in significant CTN changes and perceivable artefacts, particularly for a superficial organ close to the shield, and is not recommended for quantification CT examinations or follow-up CT examinations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。