Effect of Charpy Striker Configuration on Low- and High-Energy NIST Verification Specimens.

阅读:6
作者:Lucon Enrico, Splett Jolene
Charpy machines can be equipped with strikers having two different configurations, corresponding to an edge radius of 2 mm or 8 mm. Both striker types are covered by ASTM E23 and ISO 148-1. The effect of striker type on Charpy absorbed energy has been extensively investigated in the past, and clear evidence has been published showing that when using 8 mm strikers, absorbed energy (KV) tends to increase for specimens with KV ≥ 200 J. In this paper, we investigate how striking edge radius affects certified values and uncertainties for National Institute of Standards and Technology (NIST) low-energy and high-energy verification specimens. Test data from two low-energy and two high-energy Charpy lots, analyzed in a statistically rigorous manner, were somewhat contradictory and led to the decision to separately certify low-energy and high-energy lots for use with 2 mm and 8 mm strikers. This agrees with previous findings by other NIST researchers, who recommended individual certifications for the two strikers at all energy levels.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。