The aim of the work was to prepare a simple but reliable HPLC-UV method for the routine monitoring of mycophenolic acid (MPA). Sample preparation was based on plasma protein precipitation with acetonitrile. The isocratic separation of MPA and internal standard (IS) fenbufen was made on Supelcosil LC-CN column (150 à 4.6 mm, 5 µm) using a mobile phase: CH(3)CN:H(2)O:0.5M KH(2)PO(4):H(3)PO(4) (260:700:40:0.4, v/v). UV detection was set at 305 nm. The calibration covered the MPA concentration range: 0.1-40 µg/mL. The precision was satisfactory with RSD of 0.97-7.06% for intra-assay and of 1.92-5.15% for inter-assay. The inaccuracy was found between -5.72% and +2.96% (+15.40% at LLOQ) and between -8.82% and +5.31% (+19.00% at LLOQ) for intra- and inter-assay, respectively, fulfilling acceptance criteria. After a two-year period of successful application, the presented method has been retrospectively calibrated using the raw data disregarding the IS in the calculations. The validation and stability parameters were similar for both calculation methods. MPA concentrations were recalculated and compared in 1187 consecutive routine therapeutic drug monitoring (TDM) trough plasma samples from mycophenolate-treated patients. A high agreement (r(2) = 0.9931, p < 0.0001) of the results was found. A Bland-Altman test revealed a mean bias of -0.011 μg/mL (95% CI: -0.017; -0.005) comprising -0.14% (95% Cl: -0.39; +0.11), whereas the Passing-Bablok regression was y = 0.986x + 0.014. The presented method can be recommended as an attractive analytical tool for medical (hospital) laboratories equipped with solely basic HPLC apparatus. The procedure can be further simplified by disapplying an internal standard while maintaining appropriate precision and accuracy of measurements.
Validated Simple HPLC-UV Method for Mycophenolic Acid (MPA) Monitoring in Human Plasma. Internal Standardization: Is It Necessary?
阅读:4
作者:Kunicki PaweÅ K, Wróbel Aleksandra
| 期刊: | Molecules | 影响因子: | 4.600 |
| 时间: | 2021 | 起止号: | 2021 Nov 29; 26(23):7252 |
| doi: | 10.3390/molecules26237252 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。