The simulation of the gas flow field and electrostatic field in the photoionization detector by COMSOL was conducted based on principle investigation in the present study. Under the guidance of simulation results, structural optimization was carried out to significantly reduce the dead volume of the ionization chamber, and finally, the relationship between offset voltage and collection efficiency was obtained which led to a remarkable increase in the collection efficiency of charged ions in the photoionization detector. Then an ionization chamber with low interference and fast response was developed. Then experiment was performed with toluene as a VOCs gas under the condition of optimal gas flow rate of 50âml, UV lamp ionization energy of 10.86âeV. The results showed that the ion collection efficiency reached 91% at a bias voltage of 150âV. Moreover, a preferred linearity of 99.99% was obtained, and a ppb level of LOD can be achieved. The determination results well-fitted the relationship between offset voltage and the response value obtained in the simulation.
Development of a Novel Micro Photoionization Detector for Rapid Volatile Organic Compounds Measurement.
阅读:4
作者:Zhou Qi, Zhang Sixiang, Zhang Xu, Ma Xu, Zhou Wei
| 期刊: | Applied Bionics and Biomechanics | 影响因子: | 0.600 |
| 时间: | 2018 | 起止号: | 2018 Sep 5; 2018:5651315 |
| doi: | 10.1155/2018/5651315 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。