Degradation of Bioresorbable Mg-4Zn-1Sr Intramedullary Pins and Associated Biological Responses in Vitro and in Vivo.

阅读:4
作者:Cipriano Aaron F, Lin Jiajia, Lin Alan, Sallee Amy, Le Belinda, Cortez Alcaraz Mayra Celene, Guan Ren-Guo, Botimer Gary, Inceoğlu Serkan, Liu Huinan
This article reports the degradation and biological properties of as-drawn Mg-4Zn-1Sr (designated as ZSr41) and pure Mg (P-Mg) wires as bioresorbable intramedullary pins for bone repair. Specifically, their cytocompatibility with bone marrow derived mesenchymal stem cells (BMSCs) and degradation in vitro, and their biological effects on peri-implant tissues and in vivo degradation in rat tibiae were studied. The as-drawn ZSr41 pins showed a significantly faster degradation than P-Mg in vitro and in vivo. The in vivo average daily degradation rates of both ZSr41 and P-Mg intramedullary pins were significantly greater than their respective in vitro degradation rates, likely because the intramedullary site of implantation is highly vascularized for removal of degradation products. Importantly, the concentrations of Mg(2+), Zn(2+), and Sr(2+) ions in the BMSC culture in vitro and their concentrations in rat blood in vivo were all lower than their respective therapeutic dosages, i.e., in a safe range. Despite of rapid degradation with a complete resorption time of 8 weeks in vivo, the ZSr41 intramedullary pins showed a significant net bone growth because of stimulatory effects of the metallic ions released. However, proportionally released OH(-) ions and hydrogen gas caused adverse effects on bone marrow cells and resulted in cavities in surrounding bone. Thus, properly engineering the degradation properties of Mg-based implants is critical for harvesting the bioactivities of beneficial metallic ions, while controlling adverse reactions associated with the release of OH(-) ions and hydrogen gas. It is necessary to further optimize the alloy processing conditions and/or modify the surfaces, for example, applying coatings onto the surface, to reduce the degradation rate of ZSr41 wires for skeletal implant applications.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。