The design and construction of a high precision ambient ionization source matrix-assisted laser desorption electrospray ionization (MALDESI) are described in full detail, including a complete parts list. The computer controlled high precision motion control system and high repetition rate Explorer laser are demonstrated during MALDESI-FT-ICR analysis of peptides and proteins ranging from 1 to 17 kDa. The high stability ionization source platform described herein demonstrates both the advantages of the new MALDESI source and versatility for application to numerous desorption and ionization techniques.
Construction of a versatile high precision ambient ionization source for direct analysis and imaging.
阅读:16
作者:Sampson Jason S, Hawkridge Adam M, Muddiman David C
| 期刊: | Journal of the American Society for Mass Spectrometry | 影响因子: | 2.700 |
| 时间: | 2008 | 起止号: | 2008 Oct;19(10):1527-34 |
| doi: | 10.1016/j.jasms.2008.06.013 | ||