BACKGROUND: Impingement events, in addition to their role immediately proximate to frank dislocation, hold the potential to damage new-generation hard-on-hard bearings as a result of the relatively unforgiving nature of the materials and designs. Because of the higher stiffness and tighter design tolerances of metal-on-metal and ceramic implants, surgical positioning plausibly has become even more important. QUESTIONS/PURPOSES: We asked (1) whether, and under what cup orientation conditions, hard-on-hard impingements might challenge implant material failure strength; and (2) whether particle generation propensity at impingement and egress sites would show similar dependence on cup orientation. METHODS: Realistic computational simulations were enabled by multistage finite element analyses, addressing both global construct motion and loading, and focal stress concentrations at neck impingement and rim egress sites. The global model, validated by a cadaveric simulation in a servohydraulic hip simulator, included both hardware components and advanced anisotropic capsule characterization. Parametric computational runs explored the effect of cup orientation for both ceramic-on-ceramic and metal-on-metal bearing couples for two distinct motion sequences associated with dislocation. RESULTS: Stress concentrations from impingement increased nearly linearly with increased cup tilt and with cup anteversion. In some situations, peak values of stress approached or exceeded 1 GPa, levels challenging the yield strength of cobalt-chromium implants, and potentially the fracture strength of ceramics. The tendency for impingement events to generate debris, indexed in terms of a new scraping severity metric, showed orientation dependences similar to that for bulk material failure. CONCLUSIONS: Damage propensity arising from impingement events in hard total hip bearings is highly orientation-dependent.
Hard-on-hard total hip impingement causes extreme contact stress concentrations.
阅读:3
作者:Elkins Jacob M, O'Brien Megan K, Stroud Nicholas J, Pedersen Douglas R, Callaghan John J, Brown Thomas D
| 期刊: | Clinical Orthopaedics and Related Research | 影响因子: | 4.400 |
| 时间: | 2011 | 起止号: | 2011 Feb;469(2):454-63 |
| doi: | 10.1007/s11999-010-1632-x | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。