Steroid hormone concentrations reflect diverse physiological and pathological processes and have been recognized as valuable biomarkers for disease, with growing interest in their potential for patient stratification in precision medicine. Salivary steroid concentrations should reflect the free (biologically active) steroids in circulation, as steroids in the bloodstream passively diffuse to saliva. This allows for the direct measurement of free steroids without transporter protein-bound hormones (inactive form). However, implementation of salivary steroid quantification in larger studies remains limited by challenges associated with sample preparation. We began by reviewing the literature on relevant sample preparation approaches and identified commonly used 96-well solid phase extraction (SPE) methods for further evaluation. Three sample preparation methods were selected and assessed in terms of internal standard recovery and matrix effects, and their response was also compared using electrospray (ESI) versus UniSpray ionization (USI) liquid chromatography-tandem mass spectrometry (LC-MS/MS). We demonstrate a sensitive and rapid high-throughput method with Oasis HLB µElution SPE of 200 μL saliva in 96-well format and USI-LC-MS/MS for major steroids (testosterone, androstenedione, cortisone, cortisol, and progesterone) in saliva. The method achieved optimal recovery (77%), matrix effects (33%), and sensitivity with detection limits (MDL) ranging between 1.1 and 3.0 pg/mL and linearity (r(2)â=â0.99). Intra-plate and inter-plate coefficient of variation (CV) was below 7% and 20% using USI. USI provided a higher (2.0-2.8-fold) response than ESI and higher signal-to-noise ratio (S/N). The method was then applied to 97 authentic saliva samples (41 male and 56 female) and significant correlations between age and BMI, and androgen levels were observed in both sexes. The proposed 96-well SPE USI-LC-MS/MS method is well-suited for determining steroid hormones in saliva with potential usefulness in large-scale studies and clinical settings.
Non-invasive assessment of free steroid hormones: development of a high-throughput LC-MS/MS method for salivary steroid hormone quantification.
阅读:7
作者:Slettvoll Tove, Wahlberg Jeanette, Lood Yvonne, Josefsson Martin, Aardal Elisabeth, Salihovic Samira
| 期刊: | Analytical and Bioanalytical Chemistry | 影响因子: | 3.800 |
| 时间: | 2025 | 起止号: | 2025 Sep;417(21):4883-4894 |
| doi: | 10.1007/s00216-025-06005-9 | ||
特别声明
1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。
2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。
3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。
4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。