Phosphopeptide elution times in reversed-phase liquid chromatography.

反相液相色谱中磷酸肽的洗脱时间

阅读:10
作者:Kim Jeongkwon, Petritis Konstantinos, Shen Yufeng, Camp David G 2nd, Moore Ronald J, Smith Richard D
Elution time shifts between 33 different peptides and their corresponding phosphopeptides ranging from 4 amino acid residues to 35 amino acids in length were systematically investigated using high-resolution reversed-phase liquid chromatography (RPLC)-tandem mass spectrometry (MS/MS) analysis with trifluoroacetic acid as the ion pairing agent. Observed peptide elution time shifts for a single phosphorylation ranged from -5.28 min (for pYVPML) to +0.59 min (for HRDpSGLLDSLGR). Peptides containing a phosphotyrosine residue displayed a significant decrease in elution time following phosphorylation compared to their similar-sized peptides with phosphoserine or phosphothreonine residues. While peptide phosphorylation generally led to a decrease in the observed elution time, five peptides displayed increased elution times as a result of phosphorylation. For large peptides (> or =18 amino acids), the elution time shifts due to single phosphorylation were limited (ranging between -0.48 and +0.03 min), while the elution time shifts for small peptides (<18 amino acids) were characterized by a larger deviation (ranging between -5.28 and +0.59 min). The predictive capability for the observed RPLC elution time change due to phosphorylation has been suggested, which will aid in assigning confident phosphopeptide identifications and their subsequent confirmation.