Electrochemical biosensors have revolutionized glucose monitoring but have not yet fulfilled their promise of a low cost, direct detection replacement for genetic amplification tests such as PCR [K. Kerman, M. Kobayashi, E. Tamiya, Recent trends in electrochemical DNA biosensor technology, Meas. Sci. Technol. 15 (2004) R1-R11; A. Chaubey, B.D. Malhotra, Mediated biosensors. Biosens. Bioelectron. 17 (6-7) (2002) 441-456]. It has been anticipated that the integration of nanoscale chemical structures such as self-assembled monolayers with electrochemical biosensors would increase sensitivity by decreasing inherent system noise. We have designed a novel biosensing approach incorporating such integration and achieved rapid, ultra-low concentration sensitivities without target amplification. Raw samples are mixed with lysis buffer to allow hybridization of nucleic acid targets with anchor and signal probes before immobilizing a signaling enzyme proximate to the biosensor surface. A bias potential is subsequently applied and the secondary byproduct of a cyclic peroxidase reaction measured. Further studies have demonstrated the application of our approach in protein, clinical chemistry, and ionic assays.
Electrochemical molecular analysis without nucleic acid amplification.
无需核酸扩增的电化学分子分析
阅读:5
作者:Gau Vincent, Ma Shu-Ching, Wang Hua, Tsukuda Joni, Kibler John, Haake David A
| 期刊: | Methods | 影响因子: | 4.300 |
| 时间: | 2005 | 起止号: | 2005 Sep;37(1):73-83 |
| doi: | 10.1016/j.ymeth.2005.05.008 | ||