Flow cytometry methods used to measure leukocyte function often entail sample preparation procedures that cause artifactual cell activation. To avoid leukocyte activation by isolation techniques, some preparation methods use fluorescent markers to discriminate leukocytes from erythrocytes in whole blood. One of these markers, laser dye styryl-751(LDS-751), has been used to distinguish leukocytes by staining nucleic acid, but has been found to stain other blood cells and dead cells indiscriminately. Thus, LDS-751 may not be an appropriate reagent for leukocyte identification in whole blood. Fixing samples with formaldehydes increases cell permeability and causes surface protein cross-linking that may alter staining of both intra- and extracellular markers. The degree of this sample alteration by formaldehyde fixation, however, remains in question. In addition, little is known about flow cytometry and sample preparation methods in mouse whole blood. The purpose of this study was to determine if labeling leukocytes with a monoclonal antibody specific to leukocyte common antigen (CD45) was superior to labeling with LDS-751 and to determine the effect of sample fixation on a mouse whole blood preparation for flow cytometry. Samples were incubated with CD16/CD32 Fc receptor blocker, and either 10 microg/ml LDS-751 or phosphate buffered saline (PBS). The samples were then fixed with paraformaldehyde or diluted with PBS followed by incubation with 5 microg/ml PerCP-conjugated anti-CD45, 5 microg/ml FITC-conjugated anti-CD11b, or 80 microM dichlorofluorescein diacetate. We found that samples labeled with LDS-751 demonstrated decreased fluorescence intensity for granulocyte CD11b expression and ROS production compared to samples labeled with anti-CD45. In addition, sample fixation decreased mean fluorescence intensity in samples labeled with either LDS-751 or anti-CD45. We conclude that labeling leukocytes with monoclonal antibody CD45 in a mouse whole blood preparation is preferable, as it provides improved measurement of leukocyte indices compared to LDS-751. Also, while sample fixation prior to antibody staining caused a decrease in overall fluorescence; it can be used to successfully identify extra-cellular markers.
Comparison of sample fixation and the use of LDS-751 or anti-CD45 for leukocyte identification in mouse whole blood for flow cytometry.
比较样品固定和使用 LDS-751 或抗 CD45 对小鼠全血白细胞进行流式细胞术鉴定的效果
阅读:4
作者:Maes Melissa L, Davidson Lisa B, McDonagh Paul F, Ritter Leslie S
| 期刊: | Journal of Immunological Methods | 影响因子: | 1.600 |
| 时间: | 2007 | 起止号: | 2007 Jan 30; 319(1-2):79-86 |
| doi: | 10.1016/j.jim.2006.10.016 | 种属: | Mouse |
| 方法学: | FCM | 靶点: | CD4、CD45 |
| 研究方向: | 细胞生物学 | ||
特别声明
1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。
2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。
3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。
4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。