TaMoNbTiZr Multielement Alloy for Medical Instruments.

用于医疗器械的TaMoNbTiZr多元合金

阅读:5
作者:Mates Ileana Mariana, Geanta Victor, Manu Doina, Kelemen Hajnal, Onici Adrian Emanuel, Mirza-Rosca Julia Claudia, Voiculescu Ionelia
In this paper, a new TaMoNbTiZr multielement alloy has been designed, using chemical elements that exhibit extremely low bio-toxicity for the human body. The alloy was obtained by melting in vacuum arc remelting (VAR) equipment MRF ABJ 900 from high-purity chemical elements (99.5%) as mini-ingots having about 40 g weight each. The biocompatible alloys underwent changes in hardness after performing the annealing at 900 °C for 2 h, followed by cooling in water. The new alloy had an average hardness in the cast state of 545 HV0.5, and after heat treatment, it hardened to a value of 984 HV0.5, over 40% higher than that in the casting state, which ensures a longer working period. To use them as materials for medical instruments, their biocompatibility was highlighted through specific laboratory tests. For this, mesenchymal stem cells isolated from bone tissue and a human fibroblast cell line were cultured in vitro on the TaMoNbTiZr alloy's surface. The biocompatibility of the alloy with the biological environment was evaluated by analyzing cell viability, adhesion, and proliferation, and in parallel, the cytolysis effects manifested by the increase in lactate dehydrogenase activity in the culture media were analyzed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。