Alternative Anticoagulant Strategy to Improve the Test Sensitivity of ASTM F2888-19 Standard for Platelet and Leukocyte Count Assay.

提高 ASTM F2888-19 血小板和白细胞计数测定标准检测灵敏度的替代抗凝策略

阅读:5
作者:Patel Mehulkumar, Serna Carlos 3rd, Parrish Anna, Gupta Arjun, Jamiolkowski Megan, Lu Qijin
The ASTM F2888-19 standard for platelet and leukocyte count assay is the only standardized test method for assessing platelet and leukocyte interactions with blood-contacting device materials. This study aimed to address two limitations of the ASTM F2888-19 standard: low test sensitivity for leukocyte count and high test sample surface area to blood ratio (12 cm(2)/mL). Human blood from healthy adult donors was drawn into polypropylene tubes with either 3.2% sodium (Na) citrate or anticoagulant citrate dextrose solution A (ACDA). Immediately before starting the test, the blood was recalcified and heparinized to a concentration of 1, 1.5, or 2 U/mL and incubated with the test materials of varying thrombogenic potential at an exposure ratio of 6 or 12 cm(2)/mL for 1 h at 37°C ± 2°C in a shaking water bath. Complete blood count was measured using a hematology analyzer. The results show that both, Na-citrated blood (6 or 12 cm(2)/mL exposure ratio) and ACDA blood (6 cm(2)/mL ratio), were able to differentiate thrombogenic materials from commonly used biomaterials based on platelet count changes. The magnitudes of difference between the thrombogenic materials and biomaterials depends on heparin concentration. The test sensitivity was highest when ACDA blood, heparinized to 1 U/mL heparin, was used. Moreover, the use of ACDA blood, unlike Na-citrated blood, also allowed the assay to distinguish between the thrombogenic materials from commonly used biomaterials based on leukocyte count changes. In conclusion, the use of ACDA blood significantly increased test sensitivity of the ASTM F2888-19 test method in differentiating materials with varying thrombogenicity based on both platelet and leukocyte counts, while reducing blood exposure ratio to 6 cm(2)/mL. These findings will be used to revise the ASTM F2888 standard in the future.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。