A pilot clinical study was conducted that compared the peripheral oxygen saturation (SpO(2)) targeting performance of an automatic oxygen control system with manual oxygen control, which is the standard of care for preterm and low birth weight infants on high-flow nasal cannula (HFNC). The new oxygen control device studied was used to automatically adjust the fraction of inspired oxygen (FiO(2)) according to a desired SpO(2) target setpoint and measured feedback signals including the SpO(2) and other signals. A crossover study was designed with several endpoints including the comparison of the percentage of time that the SpO(2) was within the target range with the automatic oxygen control device versus manual oxygen control. Other metrics were also compared to assess the performance of the system including the number of bradycardia events. The pilot study included six patients that fit the inclusion criteria. The results showed that there were improvements in all of the measured outcomes considered including statistically significant improvements in the number of bradycardia events during the period when the automatic oxygen control device was used.
Clinical Evaluation of an Automatic Oxygen Control System for Premature Infants Receiving High-Flow Nasal Cannula for Respiratory Support: A Pilot Study.
对接受高流量鼻导管呼吸支持的早产儿的自动氧气控制系统进行临床评估:一项试点研究
阅读:5
作者:Hou Xuefeng, Faqeeh Akram, Amjad Ramak, Pardalos John, Fales Roger
| 期刊: | J Med Device | 影响因子: | 0.000 |
| 时间: | 2022 | 起止号: | 2022 Sep 1; 16(3):031005 |
| doi: | 10.1115/1.4054250 | ||