A simple and fast-automated method was developed and validated for the assay of promethazine hydrochloride in pharmaceutical formulations, based on the oxidation of promethazine by cerium in an acidic medium. A portable system, consisting of a programmable syringe pump connected to a potentiometer, was constructed. The developed change in potential during promethazine oxidation was monitored. The related optimum working conditions, such as supporting electrolyte concentration, cerium(IV) concentration and flow rate were optimized. The proposed method was successfully applied to pharmaceutical samples as well as synthetic ones. The obtained results were realized by the official British pharmacopoeia (BP) method and comparable results were obtained. The obtained t-value indicates no significant differences between the results of the proposed and BP methods, with the advantages of the proposed method being simple, sensitive and cost effective.
Portable system of programmable syringe pump with potentiometer for determination of promethazine in pharmaceutical applications.
用于测定药物应用中异丙嗪的带电位计的可编程注射泵便携式系统
阅读:3
作者:Saleh Tawfik A, Abulkibash A M, Ibrahim Atta E
| 期刊: | Saudi Pharmaceutical Journal | 影响因子: | 3.400 |
| 时间: | 2012 | 起止号: | 2012 Apr;20(2):155-60 |
| doi: | 10.1016/j.jsps.2011.08.005 | ||