Malaria elimination is a global public health priority. To fulfil the demands of elimination diagnostics, we have developed an interdigitated electrode sensor platform targeting the Plasmodium falciparum Histidine Rich Protein 2 (PfHRP2) protein in saliva samples. A protocol for frequency-specific PfHRP2 detection in phosphate buffered saline was developed, yielding a sensitivity of 2.5âpg/mL based on change in impedance magnitude of the sensor. This protocol was adapted and optimized for use in saliva with a sensitivity of 25âpg/mL based on change in resistance. Further validation demonstrated detection in saliva spiked with PfHRP2 from clinical isolates in 8 of 11 samples. With a turnaround time of ~2âhours, the label-free platform based on impedance sensors has the potential for miniaturization into a point-of-care diagnostic device for malaria elimination.
Ultrasensitive and label-free biosensor for the detection of Plasmodium falciparum histidine-rich protein II in saliva.
用于检测唾液中恶性疟原虫组氨酸富集蛋白 II 的超灵敏无标记生物传感器
阅读:5
作者:Soraya Gita V, Abeyrathne Chathurika D, Buffet Christelle, Huynh Duc H, Uddin Shah Mukim, Chan Jianxiong, Skafidas Efstratios, Kwan Patrick, Rogerson Stephen J
| 期刊: | Scientific Reports | 影响因子: | 3.900 |
| 时间: | 2019 | 起止号: | 2019 Nov 25; 9(1):17495 |
| doi: | 10.1038/s41598-019-53852-5 | 研究方向: | 免疫/内分泌 |