Comparison of Two Available RNA Extraction Protocols for microRNA Amplification in Serum Samples.

比较两种可用于血清样本中microRNA扩增的RNA提取方案

阅读:3
作者:Bergallo Massimiliano, Gambarino Stefano, Martino Silvana, Montin Davide, Montanari Paola, Galliano Ilaria, Tovo Pier-Angelo
BACKGROUND: microRNAs play a critical role in many biological processes such as cell proliferation and maturation, apoptosis, regulation of chronic inflammation and development of cancer. METHODS: In this study is described a protocol for the isolation of RNA from serum and subsequent determination of miRNA expression levels using TaqMan-based MGB Real-Time PCR detection. RNA was extracted using two different isolation methods including available kits RNAzol and a modified RNAzol protocol. In all cases, RNA was eluted in RNase free H2 O, kept frozen until analysis and the presence of contaminants assessed by NanoDrop spectrophotometry. RESULTS: Higher RNA quantity was observed in RNAzol (378.8 ng/μl) vs RNAzol modified protocol (226.5 ng/μl) and a better performance in terms of RNA extraction yield and purity. Subsequently, measurements of endogenous miRNAs (RNU43), cellular miRNAs (mir155 and mir146a) and EBV miRNAs (mirBART2-5p, mirBART15 and mirBART22) were performed by RT-qPCR. CONCLUSION: In contrast to the findings in terms of purity and quantity, the amplifiable RNA was more abundant using RNAzol modified protocol compared to not modified protocol.