The aims of this study were to assess damage on the surface of retrieved oxidized zirconium (OxZr) metal femoral heads, to measure surface roughness of scratches, and to evaluate the extent of surface effacement using scanning electron microscopy (SEM). Ceramic zirconia-toughened alumina heads were analyzed for comparison. OxZr femoral heads explanted for recurrent dislocation had the most severe damage (P<0.001). The median surface roughness of damaged OxZr femoral heads was 1.49μm, compared to 0.084μm for damaged ceramic heads and 0.052μm for undamaged OxZr (P<0.001). This may be of clinical concern because increased surface roughness has the potential to increase the wear of polyethylene liners articulating against these OxZr heads in THA.
Hip dislocation increases roughness of oxidized zirconium femoral heads in total hip arthroplasty: an analysis of 59 retrievals.
髋关节脱位会增加全髋关节置换术中氧化锆股骨头的粗糙度:对 59 例取出病例的分析
阅读:5
作者:Moussa Mohamed E, Esposito Christina I, Elpers Marcella E, Wright Timothy M, Padgett Douglas E
| 期刊: | Journal of Arthroplasty | 影响因子: | 3.800 |
| 时间: | 2015 | 起止号: | 2015 Apr;30(4):713-7 |
| doi: | 10.1016/j.arth.2014.10.036 | 研究方向: | 骨科研究 |