BACKGROUND: In the ongoing efforts to reduce cardiac perfusion dose (injected radioactivity) for conventional SPECT/CT systems, we performed a human observer study to confirm our clinical model observer findings that iterative reconstruction employing OSEM (ordered-subset expectation-maximization) at 25% of the full dose (quarter-dose) has a similar performance for detection of hybrid cardiac perfusion defects as FBP at full dose. METHODS: One hundred and sixty-six patients, who underwent routine rest-stress Tc-99m sestamibi cardiac perfusion SPECT/CT imaging and clinically read as normally perfused, were included in the study. Ground truth was established by the normal read and the insertion of hybrid defects. In addition to the reconstruction of the 25% of full-dose data using OSEM with attenuation (AC), scatter (SC), and spatial resolution correction (RC), FBP and OSEM (with AC, SC, and RC) both at full dose (100%) were done. Both human observer and clinical model observer confidence scores were obtained to generate receiver operating characteristics (ROC) curves in a task-based image quality assessment. RESULTS: Average human observer AUC (area under the ROC curve) values of 0.725, 0.876, and 0.890 were obtained for FBP at full dose, OSEM at 25% of full dose, and OSEM at full dose, respectively. Both OSEM strategies were significantly better than FBP with P values of 0.003 and 0.01 respectively, while no significant difference was recorded between OSEM methods (Pâ=â0.48). The clinical model observer results were 0.791, 0.822, and 0.879, respectively, for the same patient cases and processing strategies used in the human observer study. CONCLUSIONS: Cardiac perfusion SPECT/CT using OSEM reconstruction at 25% of full dose has AUCs larger than FBP and closer to those of full-dose OSEM when read by human observers, potentially replacing the higher dose studies during clinical reading.
Retrospective fractional dose reduction in Tc-99m cardiac perfusion SPECT/CT patients: A human and model observer study.
Tc-99m 心脏灌注 SPECT/CT 患者的回顾性部分剂量减少:一项人体和模型观察者研究
阅读:5
作者:Pretorius P Hendrik, Ramon Albert Juan, King Michael A, Konik Arda, Dahlberg Seth T, Parker Mathew W, Botkin Naomi F, Johnson Karen L, Yang Yongyi, Wernick Miles N
| 期刊: | Journal of Nuclear Cardiology | 影响因子: | 2.700 |
| 时间: | 2021 | 起止号: | 2021 Apr;28(2):624-637 |
| doi: | 10.1007/s12350-019-01743-7 | 种属: | Human |
| 研究方向: | 心血管 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。