Urine albumin estimation has a very high diagnostic significance. The scope of this assay widens if the method has potential for point of care deployment, while remaining truly quantitative. This study compares Proflo-U® a novel PoC device that estimates urine albumin, with standard immunoturbidity based assays. Proflo-U® is a novel PoC device which works with an android based mobile application by Bluetooth connectivity. This unique fluorescence-based urine albumin assay has been compared with the well-established gold standard Immunoturbidity based method on Beckman Coulter AU platform and also using Biosystem kit. The Proflo-U® generated data from standard recombinant human serum albumin (rHSA) and human urine spiked with rHSA has shown high correlation R (2)â~â0.99 and pâ>â0.05. This novel platform has scored 115 out of 150 among 30 parameters to determine its suitability for point of care deployment. The fluorescence-based urine albumin assay of the novel Proflo-U® platform comparable with the gold standards in the field, is easy to use, and found to be deployable as a point of care device. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s12291-024-01185-0.
Evaluation of Proflo-U® Platform: A Novel Fluorescence-Based Point of Care Device for Quantitative Estimation of Urine Albumin.
Proflo-U® 平台评估:一种用于定量测定尿白蛋白的新型基于荧光的即时检测设备
阅读:5
作者:Satyanarayana Midde Hari, Mishra Shivam, Bandyopadhyay Debapriya
| 期刊: | Indian Journal of Clinical Biochemistry | 影响因子: | 1.600 |
| 时间: | 2025 | 起止号: | 2025 Jul;40(3):445-454 |
| doi: | 10.1007/s12291-024-01185-0 | 研究方向: | 免疫/内分泌 |
特别声明
1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。
2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。
3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。
4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。