Direct mass spectrometry analysis of untreated samples of volumes as low as 0.2 µL were achieved using fast extraction and nanoESI (electrospray ionization) in a combined fashion. The analytes in dried samples on paper substrates were extracted by organic solvent in a nanoESI tube and ionized with a high voltage applied for generating a spray. The ionization source produced stable signals for different atmospheric pressure interfaces of triple quadrupole instruments. Analysis time more than 20 minutes were available with 10 µL solvent consumed for the entire analysis process. The performance in qualitative and quantitative analysis was characterized with a wide variety of samples. Limits of detection as low as 0.1 ng/mL (corresponding to an absolute amount of 0.05 pg) were obtained for analysis of atrazine in river water, thiabendazole in orange homogenate, and methamphetamine in blood.
Direct Mass Spectrometry Analysis of Untreated Samples of Ultralow Amounts Using Extraction Nano-Electrospray.
利用萃取纳米电喷雾技术对超低浓度未经处理的样品进行直接质谱分析
阅读:4
作者:Ren Yue, Liu Jiangjiang, Li Linfan, McLuckey Morgan N, Ouyang Zheng
| 期刊: | Analytical Methods | 影响因子: | 2.600 |
| 时间: | 2013 | 起止号: | 2013 Dec 7; 5(23):10 |
| doi: | 10.1039/C3AY41149D | ||