Sequential activation of ground pads reduces skin heating during radiofrequency ablation: initial in vivo porcine results

连续激活接地垫可降低射频消融过程中的皮肤加热:猪体内的初步结果

阅读:13
作者:David J Schutt, M Michael Swindle, Gorka A Bastarrika, Dieter Haemmerich

Conclusion

Sequential activation of multiple ground pads resulted in significantly lower skin temperatures during highpower RF ablation.

Methods

9 RF ablations (n=4 simultaneous, n=5 sequential) were performed with 300 W for 12 min via two internally cooled cluster electrodes placed in the gluteus maximus of domestic swine. Three ground pads placed on the animal's abdomen were activated either simultaneously, or sequentially where activation timing was adjusted to equilibrate skin temperature below each pad. Temperature rise at each pad was compared. Ablation zone dimensions were determined via MRI.

Purpose

Radiofrequency (RF) ablation is a common treatment modality for inoperable liver cancer. Skin burns below ground pads during RF ablations are increasingly prevalent, hindering the development of higher-power RF generators capable of creating larger ablation zones. Materials and

Results

Maximum temperature rise was significantly higher with simultaneous activation than with sequential activation (21.4 vs 8.1 degress C, p<0.01). Ablation zone diameters during simultaneous and sequential activation were 6.9+/-0.3 and 5.6+/-0.3, respectively.