Evaluation of Four HbA1c Point-of-Care Devices Using International Quality Targets: Are They Fit for the Purpose?

阅读:3
作者:Lenters-Westra Erna, English Emma
BACKGROUND: Point-of-care (POC) testing is becoming increasingly valuable in health care delivery, and it is important that the devices used meet the same quality criteria as main laboratory analyzers. While external quality assessment (EQA) provides a great tool for assessing quality, many POC devices are not enrolled in these schemes and standard laboratory evaluations are needed to assess performance. METHODS: The Clinical and Laboratory Standards Institute (CLSI) protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy and bias. We assessed bias using the mean of 4 certified secondary reference measurement procedures (SRMPs). RESULTS: The Afinion2™ and the Quo-Lab had CVs of ≤1.7 and ≤2.4% respectively in IFCC SI units (≤1.2 and ≤1.7% NGSP) and a bias ≤2 mmol/mol (≤0.2% NGSP) at 48 and 75 mmol/mol (6.5 and 9.0% NGSP). Sigma for the Afinion2 was 5.8 and for the Quo-Lab 4.0. Both methods passed the NGSP criteria with 2 instruments when compared with 4 individual SRMPs. The HbA1c 501 had a CV of 3.4% and 2.7% in IFCC SI units (2.1% and 1.7% NGSP) and a bias ≤2.4 mmol/mol (≤0.2% NGSP) and passed the NGSP criteria with 2 instruments compared with 4 individual SRMPs except for instrument 2 compared with the Tosoh G8. Sigma was 2.1. The A1Care had a sigma of 1.4 and failed all criteria mainly due to a high CV (6.2% and 4.1% in IFCC SI units [4.1% and 2.9% NGSP] at 48 and 75 mmol/mol [6.5 and 9.0% NGSP]). CONCLUSIONS: The analytical performance was excellent for the Afinion2 and the Quo-Lab, acceptable for the HbA1c 501 and unacceptable for the A1Care according to different used criteria, demonstrating that whilst performance is improving there are still areas for considerable improvement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。