PURPOSE: To assess dynamic tumor tracking (DTT) target localization uncertainty for in-vivo marker-based stereotactic ablative radiotherapy (SABR) treatments of the liver using electronic-portal-imaging-device (EPID) images. The Planning Target Volume (PTV) margin contribution for DTT is estimated. METHODS: Phantom and patient EPID images were acquired during non-coplanar 3DCRT-DTT delivered on a Vero4DRT linac. A chain-code algorithm was applied to detect Multileaf Collimator (MLC)-defined radiation field edges. Gold-seed markers were detected using a connected neighbor algorithm. For each EPID image, the absolute differences between the measured center-of-mass (COM) of the markers relative to the aperture-center (Tracking Error, (E(T) )) was reported in pan, tilt, and 2D-vector directions at the isocenter-plane. PHANTOM STUDY: An acrylic cube phantom implanted with gold-seed markers was irradiated with non-coplanar 3DCRT-DTT beams and EPID images collected. Patient Study: Eight liver SABR patients were treated with non-coplanar 3DCRT-DTT beams. All patients had three to four implanted gold-markers. In-vivo EPID images were analyzed. RESULTS: Phantom Study: On the 125 EPID images collected, 100% of the markers were identified. The average ± SD of E(T) were 0.24 ± 0.21, 0.47 ± 0.38, and 0.58 ± 0.37 mm in pan, tilt and 2D directions, respectively. Patient Study: Of the 1430 EPID patient images acquired, 78% had detectable markers. Over all patients, the average ± SD of E(T) was 0.33 ± 0.41 mm in pan, 0.63 ± 0.75 mm in tilt and 0.77 ± 0.80 mm in 2D directions The random 2D-error, Ï, for all patients was 0.79 mm and the systematic 2D-error, Σ, was 0.20 mm. Using the Van Herk margin formula 1.1 mm planning target margin can represent the marker based DTT uncertainty. CONCLUSIONS: Marker-based DTT uncertainty can be evaluated in-vivo on a field-by-field basis using EPID images. This information can contribute to PTV margin calculations for DTT.
In-vivo quality assurance of dynamic tumor tracking (DTT) for liver SABR using EPID images.
阅读:4
作者:Rostamzadeh Maryam, Luchka Kurt, Ma Roy, Liu Mitchell, Dunne Emma, Camborde Marie-Laure, Karan Tania, Mestrovic Ante, Bergman Alanah
| 期刊: | Journal of Applied Clinical Medical Physics | 影响因子: | 2.200 |
| 时间: | 2023 | 起止号: | 2023 Jul;24(7):e13969 |
| doi: | 10.1002/acm2.13969 | ||
特别声明
1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。
2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。
3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。
4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。
