Finite element (FE)-based studies of preoperative processes such as folding, pleating, and stent crimping with a comparison with experimental inflation tests are not yet available. Therefore, a novel workflow is presented in which residual stresses of balloon folding and pleating, as well as stent crimping, and the geometries of all contact partners were ultimately implemented in an FE code to simulate stent expansion by using an implicit solver. The numerical results demonstrate that the incorporation of residual stresses and strains experienced during the production step significantly increased the accuracy of the subsequent simulations, especially of the stent expansion model. During the preoperative processes, stresses inside the membrane and the stent material also reached a rather high level. Hence, there can be no presumption that balloon catheters or stents are undamaged before the actual surgery. The implementation of the realistic geometry, in particular the balloon tapers, and the blades of the process devices improved the simulation of the expansion mechanisms, such as dogboning, concave bending, or overexpansion of stent cells. This study shows that implicit solvers are able to precisely simulate the mentioned preoperative processes and the stent expansion procedure without a preceding manipulation of the simulation time or physical mass.
On the importance of modeling balloon folding, pleating, and stent crimping: An FE study comparing experimental inflation tests.
阅读:6
作者:Geith Markus A, Swidergal Krzysztof, Hochholdinger Bernd, Schratzenstaller Thomas G, Wagner Marcus, Holzapfel Gerhard A
| 期刊: | International Journal for Numerical Methods in Biomedical Engineering | 影响因子: | 2.400 |
| 时间: | 2019 | 起止号: | 2019 Nov;35(11):e3249 |
| doi: | 10.1002/cnm.3249 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。