All-in-One Digital Microfluidics System for Molecular Diagnosis with Loop-Mediated Isothermal Amplification.

阅读:3
作者:Hu Siyi, Jie Yuhan, Jin Kai, Zhang Yifan, Guo Tianjie, Huang Qi, Mei Qian, Ma Fuqiang, Ma Hanbin
In this study, an "all-in-one" digital microfluidics (DMF) system was developed for automatic and rapid molecular diagnosis and integrated with magnetic bead-based nucleic acid extraction, loop-mediated isothermal amplification (LAMP), and real-time optical signal monitoring. First, we performed on- and off-chip comparison experiments for the magnetic bead nucleic acid extraction module and LAMP amplification function. The extraction efficiency for the on-chip test was comparable to that of conventional off-chip methods. The processing time for the automatic on-chip workflow was only 23 min, which was less than that of the conventional methods of 28 min 45 s. Meanwhile, the number of samples used in on-chip experiments was significantly smaller than that used in off-chip experiments; only 5 µL of E. coli samples was required for nucleic acid extraction, and 1 µL of the nucleic acid template was needed for the amplification reaction. In addition, we selected SARS-CoV-2 nucleic acid reference materials for the nucleic acid detection experiment, demonstrating a limit of detection of 10 copies/µL. The proposed "all-in-one" DMF system provides an on-site "sample to answer" time of approximately 60 min, which can be a powerful tool for point-of-care molecular diagnostics.