This Letter presents an experimental study comparing the photon rate and photon economy of pulse sampling fluorescence lifetime imaging (PS-FLIm) with the conventional time-correlated single photon counting (TCSPC) technique. We found that PS-FLIm has a significantly higher photon detection rate (200â MHz) compared with TCSPC (2-8â MHz) but lower photon economy (4-5 versus 1-1.3). The main factor contributing to the lower photon economy in PS-FLIm is laser pulse variability. These results demonstrate that PS-FLIm offers 25Ã faster imaging speed than TCSPC while maintaining room light rejection in clinical settings. This makes PS-FLIm a robust technique for clinical applications.
Pulse-sampling fluorescence lifetime imaging: evaluation of photon economy.
阅读:10
作者:Zhou Xiangnan, Bec Julien, Ehrlich Katjana, Garcia Alba Alfonso, Marcu Laura
| 期刊: | Optics Letters | 影响因子: | 3.300 |
| 时间: | 2023 | 起止号: | 2023 Sep 1; 48(17):4578-4581 |
| doi: | 10.1364/OL.490096 | ||