Comparison of Temperature Programmable Split/Splitless and Cool On-Column Inlets for the Determination of Glycerol and Glycerides in Biodiesel by Gas Chromatography with Flame Ionization Detection

比较温度可编程分流/不分流进样和柱上冷进样进样在气相色谱-火焰离子化检测法测定生物柴油中甘油和甘油酯含量方面的性能

阅读:3
作者:Matthew Giardina ,James D McCurry

Abstract

The European Standard EN 14105:2011-07 is an analysis method for quantifying free glycerol and residual mono-, di- and triacylglycerides impurities in biodiesel by gas chromatography. The method specifies an "on-column injector or equivalent device" as the means of sample introduction. Cool on-column (COC) would appear to be an ideal choice, particularly for quantifying triacylglycerides, as it provides high quantitative accuracy and precision with minimal mass discrimination. However, there are a few drawbacks in using COC for this application. The relatively high concentration of the biodiesel in the prepared samples impedes solvent focusing of early eluting compounds such as glycerol, causing band broadening and shifts in retention time compared with the external calibration standards. More problematic is method robustness when using a metal retention gap. Repeated injections onto the retention gap cause the method control specification to fail within relatively few injections. As an alternative, a temperature programmable split/splitless (TPSS) inlet was investigated for performance equivalency. The results demonstrate that the TPSS yields concentration measurements indistinguishable from the COC inlet at the 95% confidence level. In addition, the robustness of the TPSS far exceeds that of the COC inlet by eliminating the performance control failure and providing solvent focusing for the early eluting peaks.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。