Discussion
LPS contamination of the placenta perfusion model could have confounding effects on experimental outcomes leading to misinterpretation of data. To circumvent LPS contamination LPS-free BSA and oxidative depyrogenation cleaning techniques should be implemented in future placental perfusion studies.
Results
TNF-α release, observed in negative control experiments, was attributed to the use of LPS-contaminated BSA as well as inadequate cleaning of the perfusion system. Once introduced in the perfusion system, LPS accumulated and created a memory effect. Oxidative but not acid-base depyrogenation effectively reduced LPS levels to concentrations that were in accordance with FDA guidelines (<0.5 EU/mL) for medical equipment redeemed appropriate for re-use.