Development and Validation of an Efficient and Highly Sensitive Enzyme-Linked Immunosorbent Assay for Alemtuzumab Quantification in Human Serum and Plasma

开发并验证一种高效、高灵敏度的酶联免疫吸附测定法,用于定量测定人血清和血浆中的阿仑单抗

阅读:7
作者:Federica R Achini-Gutzwiller, Cornelia M Jol-van der Zijde, Anja M Jansen-Hoogendijk, Arjan C Lankester, Robbert G M Bredius, Maarten J D van Tol, Dirk Jan A R Moes, Marco W Schilham

Background

Alemtuzumab is a humanized monoclonal antibody that targets the CD52 glycoprotein expressed on most lymphocytes, subsequently inducing complement-mediated and antibody-mediated cytotoxicity. Owing to its ability to induce profound immune depletion, alemtuzumab is frequently used in patients before allogeneic hematopoietic stem cell transplantation to prevent graft rejection and acute graft-versus-host disease. In this clinical context, a stable immunoassay with high sensitivity and specificity to determine alemtuzumab levels is essential for performing pharmacokinetic and pharmacodynamic analyses; however, the available

Conclusions

The developed assay provides accurate insight into alemtuzumab exposure and its effects on the clinical response to treatment, which is key to optimizing treatment strategies.

Methods

This enzyme-linked immunosorbent assay technique was developed and validated in accordance with the European Medicines Agency guidelines on bioanalytical method validation.

Results

The assay sensitivity (lower limit of quantification) is 0.5 ng·mL -1 , and the dynamic range is 0.78-25 ng·mL -1 . To accommodate quantification of peak concentration and concentrations below the lympholytic level (<0.1 mcg·mL -1 ), patients' serum samples were prediluted 20-400 times according to the expected alemtuzumab concentration. The overall within-run accuracy was between 96% and 105%, whereas overall within-run precision (coefficient of variation) was between 3% and 9%. The between-run assessment provided an overall accuracy between 86% and 95% and an overall coefficient of variation between 5% and 14%. Conclusions: The developed assay provides accurate insight into alemtuzumab exposure and its effects on the clinical response to treatment, which is key to optimizing treatment strategies.