Dual-energy computed tomography arthrography of the shoulder joint using virtual monochromatic spectral imaging: optimal dose of contrast agent and monochromatic energy level

采用虚拟单色光谱成像进行肩关节双能量计算机断层扫描造影术:最佳造影剂剂量和单色能量水平

阅读:6
作者:Chansik An, Yong-Min Chun, Sungjun Kim, Young Han Lee, Min Jeong Yun, Jin-Suck Suh, Ho-Taek Song

Conclusion

Dual-energy scan with monochromatic spectral imaging mode results in reduced image noise and beam hardening artifacts.

Methods

Dual-energy scans with monochromatic spectral imaging mode and conventional single energy scans were performed on a shoulder phantom with 10 concentrations from 0 to 210 mg/mL of iodinated contrast medium at intervals of 15 or 30 mg/mL. Image noise, tissue contrast, and beam hardening artifacts were assessed to determine the optimum dose of contrast agent and the level of monochromatic energy for DECT shoulder arthrography in terms of the lowest image noise and the least beam hardening artifacts while good tissue contrast was maintained. Material decomposition (MD) imaging for bone-iodine differentiation was qualitatively assessed. The optimized protocol was applied and evaluated in 23 patients.

Objective

To optimize the dose of contrast agent and the level of energy for dual-energy computed tomography (DECT) arthrography of the shoulder joint and to evaluate the benefits of the optimized imaging protocol. Materials and

Results

The optimal contrast dose and energy level were determined by the phantom study at 60 mg/mL and 72 keV, respectively. This optimized protocol for human study reduced the image noise and the beam-hardening artifacts by 35.9% and 44.5%, respectively. Bone-iodine differentiation by MD imaging was not affected by the iodine concentration or level of energy.