Ct values as a diagnostic tool for monitoring SARS-CoV-2 viral load using the QIAstat-Dx® Respiratory SARS-CoV-2 Panel

使用 QIAstat-Dx® Respiratory SARS-CoV-2 Panel 以 Ct 值作为监测 SARS-CoV-2 病毒载量的诊断工具

阅读:7
作者:Martí Juanola-Falgarona, Luis Peñarrubia, Sara Jiménez-Guzmán, Roberto Porco, Clàudia Congost-Teixidor, Marta Varo-Velázquez, Sonia N Rao, Gemma Pueyo, Davide Manissero, Josep Pareja

Conclusion

The SARS-CoV-2 Ct values provided by the QIAstat-Dx® Respiratory SARS-CoV-2 Panel could be used as a surrogate for viral load given the linear correlation between Ct values and viral concentration down to limit of detection. This panel allows to obtain reproducible Ct values for SARS-CoV-2 ribonucleic acid downstream of the sample collection, reducing the sample-to-Ct workflow variability. Ct values can help provide a reliable assessment and comparison of viral loads in patients when tested with the QIAstat-Dx Respiratory SARS-CoV-2 Panel.

Methods

Different lineages of SARS-CoV-2 clinical samples and the World Health Organization international standard were used to assess the linearity of the QIAstat-Dx Respiratory SARS-CoV-2 Panel. Limit of detection for the different lineages was characterized.

Results

Comparable efficiencies and linearity for all samples resulted in R2 ≥0.99, covering a dynamic range of 1,000,000-100 copies/mL for the SARS-CoV-2 assay, showing linear correlation between Ct values and viral load down to 300 copies/mL.