Rapid diagnostic method for detection of mumps virus genome by loop-mediated isothermal amplification

利用环介导等温扩增技术快速检测腮腺炎病毒基因组的诊断方法

阅读:2
作者:Takao Okafuji,Naoko Yoshida, Motoko Fujino, Yoshie Motegi, Toshiaki Ihara, Yoshinori Ota, Tsugunori Notomi, Tetsuo Nakayama

Abstract

Most mumps patients are clinically diagnosed without any virological examinations, but some diagnosed cases of mumps may be caused by other pathogens or secondary vaccine failure (SVF). To clarify these issues, a sensitive, specific, and rapid diagnostic method is required. We obtained 60 salivary swabs from 34 patients with natural infection during the course of the illness, 10 samples from patients with vaccine-associated parotitis, and 5 samples from patients with SVF. Total RNA was extracted and subjected to reverse transcription-PCR (RT-PCR) and loop-mediated isothermal amplification (LAMP) for genome amplification. We detected mumps virus RNA corresponding to 0.1 PFU by LAMP within 60 min after RNA extraction, with the same sensitivity as RT-nested PCR. Mumps virus was isolated in 30 of 33 samples within day 2, and mumps virus genome was amplified by LAMP in 32 of them. The quantity of virus titer was calculated by monitoring the time to reach the threshold of turbidity. The viral load decreased after day 3 and was lower in patients serologically diagnosed as having SVF with milder illness. Accuracy of LAMP for the detection of mumps virus genome was confirmed; furthermore, it is of benefit for calculating the viral load, which reflects disease pathogenesis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。