Kerma area product (KAP) and scatter measurements for intraoral X-ray machines using three different types of round collimation compared with rectangular beam limiter

本文对使用三种不同类型的圆形准直器和矩形光束限制器的口内X射线机进行了比释动能面积乘积(KAP)和散射测量。

阅读:3
作者:Dennise Magill,Nhan James Huu Ngo,Marc A Felice,Mel Mupparapu

Abstract

Objectives:: The objective of the study was to determine the radiation dose reduction achieved when rectangular collimation was used on various round collimators. In addition, we evaluated the tissue doses imparted to various head and neck organs. Methods:: To evaluate the variation in radiation output based on the variable geometric configurations, the kerma area product (KAP) was measured using a commercially available KAP-meter with an internal ion chamber capable of detecting both radiation dose (µGy) and the primary X-ray beam area. The KAP was measured using standard 20.4, 25.7, and 31.7 cm2 round collimators with and without rectangular X-ray field restrictors. To evaluate the potential change in patient scatter radiation dose, an adult head phantom was loaded with thin strips of gafchromic film. A full mouth X-ray series was acquired with various geometric configurations. The films were quantified using a calibration factor to yield absorbed organ doses for the eyes, thyroid, and salivary glands. Results:: With the use of rectangular collimator, the KAP for a 31.7 cm2 round collimator was reduced by up to 60% while the 20.4 cm2 round collimator elicited a reduction from up to 40%. In the organ study, results of up to 81% reduction in scatter radiation dose were observed. Conclusions:: Although, US FDA regulations allow a maximum beam size of 38.5 cm2 on the patient skin, this study suggests that the use of rectangular collimators provide clinically relevant dose reduction for patients, even when using smaller round collimation, hence the use of rectangular collimation for all intraoral radiographic procedures is highly recommended.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。