Optimizing oxytocin LC-MS/MS sensitivity by choosing the right column

通过选择合适的色谱柱来优化催产素 LC-MS/MS 灵敏度

阅读:3
作者:Mariëtte T Ackermans

Conclusions

This study shows that choosing the right column contributes to the sensitivity of the method.

Methods

The sensitivity is compared for 18 different reversed phase and 2 different HILIC columns using 2 different mobile phase compositions. Sensitivity was evaluated in terms of S/N for 1,5 pg oxytocin on column, using a scouting gradient.

Objective

Sensitivity is often an issue in bioanalytical LC-MS/MS applications. Commonly investigated parameters to improve it include additives to mobile phase, derivatization and sample-preparation. The nature of the column, however, is not frequently evaluated. Design and

Results

The measured signal to noise ranged from 55 to 1473, indicating a substantial difference in sensitivity. The most sensitive columns were the Synergi Hydro RP (reversed phase) and the Atlantis HILIC (HILIC). Conclusions: This study shows that choosing the right column contributes to the sensitivity of the method.