Conclusion
EZVent is equivalent to the commercial ventilator regarding its safety and efficacy.
Methods
Eight adult male pigs were anesthetized and randomly assigned into two equal groups: Commercial vent and EZVent group, the animals of which were ventilated using a standard commercial ventilator and EZVent, respectively. On every animal, three ventilation modes were tested, each mode for 30 min: CMV-VC, CMV-PC, and CPAP-PS modes. Vital signs, ECG, Lung Mechanics (LM), and ABG were measured before ventilation and after 30 min of ventilation of each mode. After animals' euthanasia, histological examinations of lung samples including morphometric assessment of alveolar edema, alveolar wall thickening, and the mean number of inflammatory cellular infiltrate/cm2 of lung tissue were analyzed. TNF-α and Il-6 expression and localization in lung tissue were assessed by western blot and immunohistochemistry.
Results
The vital signs, LM, ABG, morphometric analysis, and histopathological score during the different ventilation modes showed non-significant differences between the study groups. TNF-α and IL-6 were minimally expressed in the bronchiolar epithelium and the alveolar septa. Their increased expression level was insignificant.