TEG® 6s coagulation testing with a novel heparin neutralization cartridge: Technical validation and determination of normal reference ranges

采用新型肝素中和试剂盒的TEG® 6s凝血检测:技术验证和正常参考范围的确定

阅读:1

Abstract

OBJECTIVES: We sought to establish normal reference ranges (NRRs) for a novel TEG 6s cartridge (TEG 6s Citrated: K, KH, RTH, FFH [Global Hemostasis]) (Haemonetics Corporation, Boston, MA, US). METHODS: Healthy volunteers (≥18 years of age) included in this single-arm study provided single samples of whole blood. Primary end points included TEG parameters in the citrated kaolin (CK), CK with heparinase (CKH), RapidTEG with heparinase (CRTH), and functional fibrinogen with heparinase (CFFH) assays. RESULTS: Evaluable data were contributed by 164 volunteers (48.8% female; 62% White/Caucasian). The following NRRs were established: CK maximum amplitude (MA), 51.0 to 67.6 mm; CKH-MA, 51.8 to 67.9 mm; CRTH-MA, 53.0 to 68.9 mm; CFFH-MA, 15.3 to 34.4 mm; CK reaction time, 5.0 to 9.1 minutes; CKH reaction time, 4.9 to 9.4 minutes; CKH lysis 30 minutes after MA, 0% to 3.2%. Duplicate measurements demonstrated high reproducibility. CFFH-MA correlated with Clauss fibrinogen concentration (Pearson correlation coefficient, 0.74). Laboratory-based studies demonstrated maintenance of the relationship between CFFH-MA and fibrinogen up to 1344 mg/dL (hyperfibrinogenemic samples) and acceptability of heparin neutralization up to concentrations of low molecular weight and unfractionated heparin of 1.3 IU/mL and 5 IU/mL, respectively. CONCLUSIONS: This study established NRRs for the Global Hemostasis cartridge and serves as a proof of concept for the validity of results obtained using this cartridge.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。